INNOVATIONS
Year : 2020  |  Volume : 152  |  Issue : 7  |  Page : 249-250

Big solution with small intervention


Department of Anaesthesiology & Critical Care, Pt. B.D.Sharma Post Graduate Institute of Medical Sciences, Rohtak 124 001, Haryana, India

Date of Submission20-Nov-2019
Date of Web Publication25-May-2021

Correspondence Address:
Prashant Kumar
Department of Anaesthesiology & Critical Care, Pt. B.D.Sharma Post Graduate Institute of Medical Sciences, Rohtak 124 001, Haryana
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijmr.IJMR_2267_19

Rights and Permissions

How to cite this article:
Kumar P, Sharma J. Big solution with small intervention. Indian J Med Res 2020;152, Suppl S1:249-50

How to cite this URL:
Kumar P, Sharma J. Big solution with small intervention. Indian J Med Res [serial online] 2020 [cited 2021 Jul 29];152, Suppl S1:249-50. Available from: https://www.ijmr.org.in/text.asp?2020/152/7/249/316803

Many supraglottic airway devices are available for use in multiple patients, with autoclaving as the permitted method of sterilization. However, these can be damaged from bite, from contact with sharp instrument, and during cleaning. Such a damage to the inflation line of ProSeal laryngeal mask airway (PLMA) was encountered at the department of Anaesthesiology & Critical Care, Pt. B.D. Sharma, Post Graduate Institute of Medical Sciences, Rohtak, Haryana, India. [Figure 1]A. Size 6 Fr feeding tube was used as a conduit [Figure 1]B, further secured with resin sealant [Figure 1]C provided adequate [Figure 1]D working, thus avoiding wastage of otherwise useful device. The seal was checked as per the manufacturers' guideline of positive and negative inflation test. To avoid any chance of compromise of patient safety, the repaired PLMA was taken to the skill laboratory for students to learn and practice with the device. Considering high cost of the device, a repair is worth attempting. Use of autoclavable material will further ensure the integrity and without breaching patient safety within the manufacturer's recommendations. Big Solutions can be achieved through small interventions, especially in a resource-limited country. Further, these repaired devices can easily be used for teaching and training to avoid possible compromise of patient safety.
Figure 1: (A) PLMA with damage to inflation line. (B) Size 6 Fr feeding tube. (C) Feeding tube inserted within damaged ends of inflation line, secured with resin sealant (arrow). (D) Well-inflated functional PLMA after intervention.

Click here to view


Conflicts of Interest: None.


    Figures

  [Figure 1]



 

Top
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Article Figures

 Article Access Statistics
    Viewed120    
    Printed0    
    Emailed0    
    PDF Downloaded23    
    Comments [Add]    

Recommend this journal